Remote Control Commands
R&S
®
SMW200A
1250User Manual 1175.6632.02 ─ 30
Parameters:
<Gain> float
Range: -1 to 1
Increment: 1E-6
*RST: 1
Example:
TEST:BB:GENerator:GAIN 0.5
// amplifies the test signal by a factor of 0.5
Manual operation: See "Gain" on page 1287
:TEST:BB:GENerator:SOURce
<IqSource>
Selects the test signal source.
Parameters:
<IqSource> SINE | CONStant | ARB
SINE
Generates a sine waveform. To determine the frequency, use
command :TEST:BB:GENerator:FREQuency on page 1249.
CONStant
Uses a constant I/Q test signal.
ARB
Selects a signal generated generated by the ARB.
*RST: SINE
Example:
:TEST:BB:GENerator:SOURce SINE
Manual operation: See "IQ Source" on page 1286
:TEST:BB:GENerator:STATe <State>
Starts the test generator.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
TEST:BB:GENerator:STATe ON
// activates the test generator
Manual operation: See "State" on page 1286
:TEST<hw>:DIRect:BLOCk:READ? <Board>, <SubAddr>, <Addr>[, <Len>]
Direct mode for querying the blocks of the hardware assemblies directly.
The respective hardware assembly responds directly to the command; any safety
mechanisms are bypassed. This function is only available via remote control.
Query parameters:
<Board> string
TEST Subsystem