Configuring the Internal Baseband Source
R&S
®
SMW200A
405User Manual 1175.6632.02 ─ 30
"Phase"
Sets the starting phase of a carrier.
If you use the Table Setup Assistant settings to fill the individual car-
rier phases, the phase of a carrier is calculated as follows:
"Phase Start" + n*"Phase Step".
"Accept"
Transfers the settings in the carrier table into the instrument.
Remote command:
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:STATe on page 1078
[:SOURce<hw>]:BB:MCCW:CARRier:STATe on page 1078
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:POWer on page 1080
[:SOURce<hw>]:BB:MCCW:CARRier:POWer on page 1080
[:SOURce<hw>]:BB:MCCW:CARRier:LIST:PHASe on page 1079
[:SOURce<hw>]:BB:MCCW:CARRier:PHASe on page 1079
5.12.3.3 Carrier Graph
Access:
► Select "Baseband > Multi-Carrier CW > Carrier Graph".
The "Carrier Graph" tab is a graphical representation of the current carrier configura-
tion.
The carriers are on the X-axis and the colored bars represent those carriers which are
in the active state. Power is on the Y-axis, and the height of the bars corresponds to
the chosen power of each individual carrier.
5.12.3.4 Trigger, Marker and Clock Setting
These tabs provide standard settings.
Generating Multi-Carrier Continuous Wave Signal