Configuring the RF Signal
R&S
®
SMW200A
590User Manual 1175.6632.02 ─ 30
Sensor Settings
One tab per sensor provides the corresponding setting parameters.
State ← Sensor Settings
Activates level measurement.
Remote command:
:INITiate<hw>[:POWer]:CONTinuous on page 894
To query the availability of a sensor at a given connector, use the command :
SENSe<ch>[:POWer]:STATus[:DEVice]? on page 902.
Zero ← Sensor Settings
Activates the auto zeroing.
For details, see "About zeroing" on page 587.
Remote command:
:SENSe<ch>[:POWer]:ZERO on page 903
Display ← Sensor Settings
Sets the display mode for power readings.
Permanent ← Display ← Sensor Settings
Activates the permanent indication of the power measurement result in the task bar.
The instrument shows the type of sensor, the used connector, the measurement
source and, if set, the offset.
You can activate the permanent display for several sensors.
Remote command:
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe on page 897
Display ← Display ← Sensor Settings
Sets the display of results on mean or peak power.
Remote command:
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority on page 897
Use Frequency Of ← Sensor Settings
Selects the source for measurement.
"RF"
Assigns one of the RF signals of the R&S SMW.
The R&S SMW transfers the RF frequency and level settings to the
R&S power sensor automatically. Thus you achieve power readings
of high accuracy, irrespective from the connected sensor type.
You can also allocate two sensors in one path, for example to mea-
sure the power before and after the DUT.