Multi-Instrument Setups
R&S
®
SMW200A
661User Manual 1175.6632.02 ─ 30
Setup Overview
Access:
Select "RF Ports Alignment > Info" > "Setup Overview".
Figure 10-8: RF Ports Setup Info > Setup Overview
This dialog displays the wiring diagram of the setup. The displayed information
resembles the setup displayed in the "RF Ports Alignment" dialog.
See "Test Setup" on page 653.
Ref Configuration > Signal Source..............................................................................661
LO Configuration > Signal Source...............................................................................661
LO Configuration > Wiring Constellation.....................................................................662
Test Setup................................................................................................................... 662
Ref Configuration > Signal Source
Indicates if the current instrument uses its own or an external reference signal.
"Primary"
The instrument uses its reference signal and provides it to the other
instruments.
"External"
The instrument uses an external reference frequency source, for
example from other R&S SMW (the one with "Ref Configuration" >
"Signal Source = Primary") or from R&S
®
SMA100B.
Remote command:
:SCONfiguration:RFALignment:SETup:INFO:WIRing:REF:SOURce?
on page 943
LO Configuration > Signal Source
Indicates if the current instrument uses its own or an external LO signal.
Generating Phase-Coherent Synchronous and Aligned Signals