Remote Control Commands
R&S
®
SMW200A
962User Manual 1175.6632.02 ─ 30
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Manual operation: See "DIG IQ Auto Setting" on page 164
[:SOURce<hw>]:BBIN:MPERiod <MPeriod>
For [:SOURce<hw>]:BBIN:DIGital:SOURce CODER1|CODER2
Sets the recording duration for measuring the baseband input signal by executed [:
SOURce<hw>]:BBIN:ALEVel:EXECute.
Parameters:
<MPeriod> integer
Range: 1 to 32
*RST: 2
Default unit: s
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Manual operation: See "Measurement Period" on page 164
[:SOURce<hw>]:BBIN:ALEVel:EXECute
For [:SOURce<hw>]:BBIN:DIGital:SOURce CODER1|CODER2
Starts measuring the input signal. The measurement estimates the crest factor, peak
and RMS level.
Example:
See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Usage: Event
Manual operation: See "Auto Level Set" on page 164
[:SOURce<hw>]:BBIN:POWer:CFACtor <CFactor>
Sets the crest factor of the external baseband signal.
Parameters:
<CFactor> float
Range: 0 to 30
Increment: 0.01
*RST: 0
Default unit: dB
Example: See Example "Applying an external digital baseband signal (DIG
I/Q)" on page 955.
Manual operation: See "Crest Factor" on page 164