Applications
R&S
®
RTM3000
199User Manual 1335.9090.02 ─ 09
Frequency
Defines the center frequency for the capture range if "R-Marker" = "Range".
Remote command:
SPECtrum:MARKer:REFerence:SETup:FREQuency on page 552
Span
Defines frequency range around the center frequency if "R-Marker" = "Range".
Remote command:
SPECtrum:MARKer:REFerence:SETup:SPAN on page 553
Center to Screen
Sets the center frequency to the center of the display.
Remote command:
SPECtrum:MARKer:REFerence:SETup:CSCReen on page 552
Center to Highest Peak
Sets the highest detected peak to the center of the display.
Remote command:
SPECtrum:MARKer:REFerence:SETup:CMPeak on page 552
Result Values
Defines the values that are shown in the peak list. You can use absolute values, or
delta values in comparison to the reference marker.
"Absolute"
Frequency and magnitude in absolute values.
"R-Marker Frequency"
Frequency relative to the reference marker frequency, magnitude in
absolute values.
"R-Marker Level"
Frequency in absolute values, magnitude relative to the level of the
reference marker.
Spectrum analysis and spectrogram (option R&S RTM-K37)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz RTM3K-COM2 and is the answer not in the manual?

Rohde & Schwarz RTM3K-COM2 Specifications

General IconGeneral
BrandRohde & Schwarz
ModelRTM3K-COM2
CategoryTest Equipment
LanguageEnglish

Related product manuals