Loading...
Signal generation (option R&S RTM-B6)
R&S
®
RTM3000
406User Manual 1335.9090.02 ─ 09
For the "Exponential" function, you can choose between a rising or falling exponential.
For the "Ramp" function, you can set a positive or negative polarity.
Remote command:
WGENerator:FUNCtion:EXPonential:POLarity on page 781
WGENerator:FUNCtion:RAMP:POLarity on page 781
Edge Time
Sets the pulse edge time.
Remote command:
WGENerator:FUNCtion:PULSe:ETIMe on page 780
Duty Cycle
Sets the duty cycle for the pulse function. The duty cycle expresses for what percent-
age of the period, the signal state is high.
Remote command:
WGENerator:FUNCtion:PULSe:DCYCle on page 780
Symmetriy
Sets the symmetry for the triangle waveform. 50% defines symmetric triangles. Values
<50% define triangles with steeper rising edge leaned to the left. Values >50% define
triangles with steeper falling edge leaned to the right.
Remote command:
WGENerator:TRIangle:SYMMetry on page 780
Arbitrary Setup
Opens a menu to configure the arbitrary waveform.
See Chapter 16.1.6, "Arbitrary setup settings", on page 411.
Sweep
Opens a menu to configure the sweep.
See Chapter 16.1.3, "Sweep settings", on page 407.
Modulation
Opens a menu to configure the modulation.
See Chapter 16.1.4, "Modulation settings", on page 408.
Invert
Inverts the waveform at the offset level.
Load
Select the user load, the load of the DUT at its connection. You can select either a
"50Ω" or a "High-Z" (high input impedance) load.
Remote command:
WGENerator:OUTPut:LOAD on page 781