Waveform setup
R&S
®
RTM3000
67User Manual 1335.9090.02 ─ 09
Current probes R&S RT-ZCxx
The current probes R&S RT-ZCxx have BNC connectors. The setup is described in
Chapter 5.4.2, "Probe settings for probes with BNC connector", on page 62. Demag-
netizing and zero adjustment is done on the probe, see the probe's User Manual for
details. Make sure to demagnetize and adjust the probe before taking measurements.
Current probes R&S RT-ZCxxB
Current probes R&S RT-ZCxxB have a Rohde & Schwarz probe interface; they are
powered and remotely controlled by the oscilloscope.
When the probe is connected, demagnetization is performed automatically.
Access: [Ch <n>] > "Probe" (scroll down); or short menu > "Probe"
Zero Adjust
Sets the waveform to zero position. After demagnetizing, always carry out a zero
adjustment. The setting corresponds to Zero Adjust in the channel menu but the value
is set in %.
Remote command:
PROBe<m>:SETup:ADVanced:ZADJust on page 448
Save Zero Adjust
Saves the "Zero Adjust" value in the probe box. If you connect the probe to another
channel or to another oscilloscope with Rohde & Schwarz probe interface, the value is
read out again.
Remote command:
PROBe<m>:SETup:ADVanced:STPRobe on page 447
Degauss
Demagnetizes the core if it has been magnetized by switching the power on and off, or
by an excessive input. Always carry out demagnetizing before measurement. The
demagnetizing process takes about one second. During demagnetizing, a demagnetiz-
ing waveform is present at the output.