Instrument setup
R&S
®
RTP
117User Manual 1337.9952.02 ─ 12
Camera hardkey action
The Camera key on the left side of the display is a shortcut key that initiates an associ-
ated action.
You can assign one of the following actions:
Save a screenshot
Open screenshot setup
Save a report
Open report setup
Configure the settings for the selected action.
Screenshots: "Save/Recall" key > "Save" tab > "Screenhot", see Chapter 12.3.1,
"Screenshot settings", on page 503.
Report: "Save/Recall" key > "Save" tab > "Report", see Chapter 12.4.1, "Report
settings", on page 507.
Quick Action
The Quick Action key on the left side of the display is a shortcut key that initiates an
associated action. To test the setup, tap "Execute".
You can assign one of the following actions to the Quick Action key:
"Application"
Starts an external application. Select the path of the application exe-
cutable, additional parameters, and the working directory as in a Win-
dows shortcut definition.
"Graphical
Recall"
Opens the "Load saveset" window to select and load instrument set-
tings that were previously stored in a saveset. See also:"Graphical
Saveset" on page 76 .
"Clear all"
Deletes the all measurement results, waveforms, and the history. See
also: "Clear" on page 76
Remote command:
QACTion:MODE on page 1182
QACTion:EXECute on page 1182
To run an application:
QACTion:PATH on page 1182
QACTion:PARameters on page 1183
QACTion:WDIRectory on page 1183
Frontpanel settings