Remote control commands
R&S
®
RTP
1171User Manual 1337.9952.02 ─ 12
DISPlay:SIGNal:LABel:REMove <LabelID>, <Source>
Deletes the specifies waveform label.
Setting parameters:
<LabelID> String with the label identifier.
<Source> All waveforms that can be displayed, see DISPlay:SIGNal:
LABel:ADD
Example:
DISPlay:SIGNal:LABel:REMove 'Label1', C1W1
Usage: Setting only
Asynchronous command
DISPlay:SIGNal:LABel:SHOW <Source>, <State>
DISPlay:SIGNal:LABel:SHOW? <Source>
Selects the source for the label display and enables it.
Parameters:
<State> ON | OFF
Instrument setup