Remote control commands
R&S
®
RTP
1202User Manual 1337.9952.02 ─ 12
Parameters:
<Offset> Negative values move the waveform up, positive values move it
down.
Range: Depends on attenuation factors, input coupling, and
the offset compensation range of active probes.
The nominal offset range for 1:1 attenuation and
probe offset compensation = 0 is specified in the
data sheet.
Increment: Depends on vertical and probe settings
*RST: 0
Default unit: V
Usage: Asynchronous command
CHANnel<m>:INVert <InvertChannel>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the volt-
age values of all signal components against the ground level. If the inverted channel is
the trigger source, the instrument triggers on the inverted signal.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<InvertChannel> ON | OFF
*RST: OFF
Usage: Asynchronous command
Firmware/software: FW 3.30
CHANnel<m>:BANDwidth <BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<BandwidthLimit> FULL
FULL
Use full bandwidth.
*RST: FULL
Usage: Asynchronous command
Acquisition and setup