Remote control commands
R&S
®
RTP
1232User Manual 1337.9952.02 ─ 12
Usage: Query only
Asynchronous command
ZVC:Z<m>:I<n>:IMPedance <MeasImp>
Sets the current impedance of the probe channel for power calculations and measure-
ments.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<MeasImp> Range: 1 to 100E+3
Increment: 1
*RST: 50
Default unit:
Ω
Usage: Asynchronous command
ZVC:Z<m>:I<n>:OFFSet <VerticalOffset>
Sets the offset current for the current channel.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<VerticalOffset> Range: -1 to 1
Increment: 0.01
*RST: 0
Default unit: V
Usage: Asynchronous command
ZVC:RESCoupled <CoupleToAnalog>
Sets the resolution of all R&S RT-ZVC channels.
Parameters:
<CoupleToAnalog> ON | OFF
ON
The resolution of the analog channels is applied to R&S RT-ZVC
channels. The signal is automatically interpolated or decimated
to get the analog resolution.
OFF
The resolution of R&S RT-ZVC channels is set in a way so that
the record length of the waveforms is minimum 1000 samples.
*RST: OFF
Acquisition and setup