EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1231

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1231User Manual 1337.9952.02 ─ 12
ZVC:Z<m>:V<n>[:STATe]..............................................................................................1239
ZVC:Z<m>:ID:NAME?.................................................................................................. 1239
ZVC:Z<m>:ID:PARTnumber?........................................................................................ 1240
ZVC:Z<m>:ID:SRNumber?............................................................................................1240
ZVC:Z<m>:ID:SWVersion?............................................................................................1240
ZVC:Z<m>:I<n>:ZERComp:DETect................................................................................1240
ZVC:Z<m>:I<n>:ZERComp:USE................................................................................... 1241
ZVC:Z<m>:I<n>:ZERComp:STATe?...............................................................................1241
ZVC:BANDwidth <Bandwidth>
Sets the bandwidth limit of the probe. The bandwidth specifies the maximum frequency
at which a purely sinusoidal signal is still transferred at 89 % (0.1 dB) of its amplitude.
Parameters:
<Bandwidth> Range: 5000 to 1E+6
Increment: 5000
*RST: 1E+6
Default unit: Hz
Usage: Asynchronous command
ZVC:TYPE <DecimationMode>
Sets the decimation mode for the R&S RT-ZVC probe. Decimation reduces the data
stream of the ADC to a stream of waveform points with lower sample rate and a less
precise time resolution.
Parameters:
<DecimationMode> SAMPle | PDETect | HRESolution
*RST: SAMPle
Usage: Asynchronous command
ZVC:Z<m>:I<n>:BANDwidth?
Queries the bandwidth of the current channel. You can set the probe bandwidth with
ZVC:BANDwidth.
The bandwidth of some current channels is restricted to 300KHz due to their vertical
settings.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<Bandwidth> Range: 5000 to 1E+6
Increment: 5000
*RST: 1E+6
Default unit: Hz
Acquisition and setup

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals