Remote control commands
R&S
®
RTP
1310User Manual 1337.9952.02 ─ 12
TRIGger<m>:SEQuence:RESet:EVENt..........................................................................1312
TRIGger<m>:SEQuence:RESet:TIMeout[:ENABle]..........................................................1312
TRIGger<m>:SEQuence:RESet:TIMeout:TIME............................................................... 1312
TRIGger<m>:EVENt <Class>
Selects, if you want to trigger on a single event, or on a series of events.
Suffix:
<m>
.
1..3
Parameters:
<Class> SINGle | SEQuence
*RST: SINGle
Usage: Asynchronous command
TRIGger<m>:SEQuence:TYPE
<Sequence>
Selects the type of the sequence. All trigger sequences require that input channels
CH1...4 are set as trigger sources for all conditions.
See also: Chapter 6.8, "Sequence", on page 246.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Sequence> ABR | AZ | ABRZ | AORB | AORBZ
ABR
Triggers if all conditions of A- and B-events, as well as additional
delay and reset timeout or R-event (reset) conditions are fulfil-
led.
AZ
Triggers if the conditions of the A-event and the zone trigger are
fulfilled. Additionally, a holdoff condition can be set.
Requires option R&S RTP-K19.
ABRZ
Triggers if all conditions of A- and B-events, reset timeout or R-
event (reset) conditions are fulfilled for the defined zone.
Requires option R&S RTP-K19.
AORB
Triggers if the conditions of A-event or the conditions of the B-
event are fulfilled.
AORBZ
Triggers if the conditions of A-event or the conditions of the B-
event are fulfilled for the defined zone. Additionally, a holdoff
condition can be set.
Requires option R&S RTP-K19.
*RST:
ABR
Trigger