Remote control commands
R&S
®
RTP
1330User Manual 1337.9952.02 ─ 12
REFCurve<m>:RESTore
Resets the time scale and the reference point to the original values of the reference
waveform.
Suffix:
<m>
.
1..4
Reference waveform
Usage: Event
Asynchronous command
REFCurve<m>:VMODe <VerticalMode>
Selects the coupling of vertical settings.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<VerticalMode> COUPled | INDependent
COUPled
Vertical position and scale of the source are used.
INDependent
Scaling and position can be set specific to the reference wave-
form.
*RST: INDependent
Usage: Asynchronous command
REFCurve<m>:SCALe <VerticalScale>
Sets the scale factor for the reference waveform if vertical scaling is set to "Independ-
ent" (see REFCurve<m>:VMODe on page 1330).
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<VerticalScale> Range: 1E-15 to 100E+24
Increment: 10E-6
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
REFCurve<m>:POSition <VertPosi>
Moves the reference waveform and its horizontal axis up or down in the diagram, if ver-
tical scaling is set to "Independent" (see REFCurve<m>:VMODe on page 1330).