EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1411

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1411User Manual 1337.9952.02 ─ 12
MEASurement<m>:MULTiple <MultiMeas>
The measurement is performed repeatedly if the measured parameter occurs several
times inside the acquisition or defined gate.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<MultiMeas> ON | OFF
*RST: OFF
Usage: Asynchronous command
MEASurement<m>:MNOMeas <MaxMeasPerAcq>
Sets the maximum number of measurements per acquisition if repeated measurement
is enabled (MEASurement<m>:MULTiple is ON).
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<MaxMeasPerAcq> Range: 2 to 1000000
Increment: 1
*RST: 1000
Usage: Asynchronous command
MEASurement<m>:STATistics:HISTogram <ShowHistogram>
Displays a histogram of the statistical results. Enabling the histogram enables also the
calculation and display of statistics for the measurement results if statistics were dis-
abled. the histogram shows the cumulative occurence distribution of mean measure-
ment results in a graphic.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ShowHistogram> ON | OFF
*RST: OFF
Usage: Asynchronous command