Remote control commands
R&S
®
RTP
1436User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:ULEVel 0.01
Sets the upper reference level for Ch1Wfm1 to 10 mV. C1W1
corresponds to suffix number 2.
Example: See: "Manual reference level definition using absolute values"
on page 1129
Usage: Asynchronous command
REFLevel<m>:ABSolute:LLEVel
<Value>
The lower reference level, required e.g. to determine a fall.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:LLEVel 0.001
Sets the lower reference level for Ch1Wfm1 to 1 mV. C1W1 cor-
responds to suffix number 2.
Example: See: "Manual reference level definition using absolute values"
on page 1129
Usage: Asynchronous command
REFLevel<m>:RELative:UPPer <Value>
Sets the upper relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value> Percentage of the high signal level.
Automatic measurements