Remote control commands
R&S
®
RTP
1437User Manual 1337.9952.02 ─ 12
Example:
REFLevel8:RELative:LOWer 85
Sets the upper reference level for Ch3Wfm1 to 85 %. Ch3Wfm1
corresponds to suffix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1129
Usage: Asynchronous command
REFLevel<m>:RELative:MIDDle <Value>
Sets the middle relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value> Percentage of the high signal level.
Example:
REFLevel8:RELative:MIDDle 50
Sets the middle reference level for Ch3Wfm1 to 50 %. Ch3Wfm1
corresponds to suffix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1129
Usage: Asynchronous command
REFLevel<m>:RELative:LOWer <Value>
Sets the lower relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value> Percentage of the high signal level.
Example:
REFLevel8:RELative:LOWer 15
Sets the lower reference level for Ch3Wfm1 to 15 %. Ch3Wfm1
corresponds to suffix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1129
Usage: Asynchronous command
Automatic measurements