Remote control commands
R&S
®
RTP
1438User Manual 1337.9952.02 ─ 12
22.12.14.4 Hysteresis
REFLevel<m>:RELative:HYSTeresis <Value>
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value>
Usage: Asynchronous command
22.12.14.5 Tube
REFLevel<m>:RELative:OTUBe....................................................................................1438
REFLevel<m>:RELative:ITUBe..................................................................................... 1438
REFLevel<m>:ABSolute:TOTube...................................................................................1439
REFLevel<m>:ABSolute:TITube.................................................................................... 1439
REFLevel<m>:ABSolute:BITube....................................................................................1439
REFLevel<m>:ABSolute:BOTube.................................................................................. 1440
REFLevel<m>:RELative:OTUBe <Value>
Defines a percentage of the signal level by which the absolute signal level may be
larger than the high signal level or lower than the low signal level to be considered high
or low, respectively.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see Chapter 22.4.1,
"Waveform suffix", on page 1141.
Parameters:
<Value>
Usage: Asynchronous command
REFLevel<m>:RELative:ITUBe <Value>
Defines a percentage of the signal level by which the absolute signal level may be
higher than the low signal level or lower than the high signal level to be considered low
or high, respectively.
Automatic measurements