Remote control commands
R&S
®
RTP
1460User Manual 1337.9952.02 ─ 12
MTESt:REMove <MaskTestName>
Deletes the mask test definition with the specified name.
Setting parameters:
<MaskTestName> String with the name of the mask test
Usage: Setting only
Asynchronous command
MTESt[:STATe] <MaskTestName>,<State>
MTESt[:STATe]? <MaskTestName>
Activates and deactivates the mask test. If the acquisition is running, the test starts
immediately. Otherwise, the test starts when acquisition is started.
The testing is stopped when acquisition is stopped, also due to the MTESt:
ONViolation:STOP command, or if MASK[:STATe] is set to "OFF".
The coommand needs *OPC command synchronisation.
Parameters:
<State> ON | OFF
*RST: OFF
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Example:
MTEST:STAT 'MyMask', ON; *OPC?
See Chapter 22.3.3.1, "Creating a user mask", on page 1132
Usage: Asynchronous command
MTESt:RST
Clears all totals and results in all "Mask Test" result boxes.
Usage:
Event
Asynchronous command
MTESt:SOURce <MaskTestName>,<Source>
MTESt:SOURce? <MaskTestName>
Selects the waveform to be tested against the mask.