Remote control commands
R&S
®
RTP
1489User Manual 1337.9952.02 ─ 12
Parameters:
<Polarity> POSitive | NEGative | EITHer
See Chapter 22.4.4, "Polarity parameter", on page 1147.
*RST:  POSitive
Parameters for setting and query: 
<SearchName> Search definition
Usage:  Asynchronous command
SEARch:TRIGger:RUNT:RANGe <SearchName>,<Mode>
SEARch:TRIGger:RUNT:RANGe? <SearchName>
Selects how the time limit of the runt pulse is defined based on the runt width and delta
(see SEARch:TRIGger:RUNT:WIDTh on page 1489 and SEARch:TRIGger:RUNT:
DELTa on page 1488).
Parameters:
<Mode> ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
Triggers on runts longer than the given "Runt width".
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "±Delta".
OUTSide
Triggers if the runt length is outside a given time range. The
range definition is the same as for "Within" range.
*RST: 
ANY
Parameters for setting and query: 
<SearchName> Search definition
Usage:  Asynchronous command
SEARch:TRIGger:RUNT:WIDTh <SearchName>,<Width>
SEARch:TRIGger:RUNT:WIDTh? <SearchName>
For the ranges "Shorter" and "Longer", the runt width defines the maximum and mini-
mum pulse width, respectively.
For the ranges "Within" and "Outside", the runt width defines the center of a range
which is defined by "±Delta".
The range is defined using SEARch:TRIGger:RUNT:RANGe on page 1489.
Search