EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 1594

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1594User Manual 1337.9952.02 ─ 12
Return values:
<FrameData> Comma-separated sequence of integer values (N, L1, R1,…,
LN, RN). N is the number of word pairs in the frame, and {L1,R1}
…{LN,RN} are the value pairs. The values Lx and Rx are associ-
ated with the MOSI and the MISO channel, respectively. If a
channel is disabled, an empty value is returned.
Example:
BUS:SPI:FRAMe3:DATA?
<-- 2,10,108,35,70 (MOSI+MISO)
2,10,,35, (MOSI only)
2,,108,,70 (MISO only)
Usage: Query only
BUS<m>:SPI:FCOunt?
BUS<m>:SPI:FRAMe<n>:COUNt?
Returns the number of decoded frames.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Return values:
<Count> Total number of decoded frames.
Usage: Query only
BUS<m>:SPI:FRAMe<n>:BITRate?
Returns the primary bit rate.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<PrimaryBitRate> Range: 0 to 100000000000
Increment: 1
*RST: 0
Default unit: bps
Usage: Query only
BUS<m>:SPI:FRAMe<n>:STATus?
Returns the overall state of the specified frame.