Remote control commands
R&S
®
RTP
1599User Manual 1337.9952.02 ─ 12
SEARch:TRIGger:SPI:DPOSition...................................................................................1600
SEARch:TRIGger:SPI:DPTO.........................................................................................1601
SEARch:TRIGger:SPI:PALignment................................................................................ 1601
SEARch:TRIGger:SPI:MODE <SearchName>,<Type>
SEARch:TRIGger:SPI:MODE? <SearchName>
Sets the event to be searched for.
Parameters:
<Type> SSACtive | TIMeout | MOSI | MISO | MOMI
SSACtive
Searches for the start of the frame when slave select signal SS
changes to the active state. This type is available if the slave
select line is configured in the bus setup, and BUS<m>:SPI:
FRCondition is SS.
TIMeout
Searches for the start of the frame when the clock idle time
exceeds the timeout. This type is available if the slave select line
is configured in the bus setup, and BUS<m>:SPI:
FRCondition is CLKTimeout.
MOSI | MISO
Searches for a specified data pattern expected on the MOSI line
or on the MISO line, respectively.
MOMI
Searches in parallel for specified data patterns expected on the
MOSI and MISO lines.
*RST:
SSACtive
Parameters for setting and query:
<SearchName>
Usage: Asynchronous command
Firmware/software: FW 3.30
SEARch:TRIGger:SPI:FCONdition <SearchName>,<DataOperator>
SEARch:TRIGger:SPI:FCONdition? <SearchName>
Selects the operator for the data pattern: equal or not equal.
Parameters:
<DataOperator> EQUal | NEQual
*RST: EQUal
Parameters for setting and query:
<SearchName>
Firmware/software: FW 3.30