Remote control commands
R&S
®
RTP
1671User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<ChecksumValue> To set the value format, use FORMat:BPATtern.
The values below range, increment and reset are decimal
values.
Range: 0 to 255
*RST: 0
Usage: Query only
BUS<m>:LIN:FRAMe<n>:BITRate?
Returns the primary bit rate.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<PrimaryBitRate> Range: 0 to 100000000000
Increment: 1
*RST: 0
Default unit: bps
Usage: Query only
BUS<m>:LIN:FRAMe<n>:BYTE<o>:STATe?
Returns the state of the specified byte.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
<o> *
Selects the byte number.
Return values:
<ByteState> OK | STERror | SPERror | UVAL | NOEXists | INSufficient
STERror: start error
SPERror: stop error
UVAL: unexpected value
Protocols