Remote control commands
R&S
®
RTP
1887User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..3
Parameters:
<CRCOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding TRIGger<m>:MPHY:CRC:MIN
command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MPHY:CRC:MIN and
TRIGger<m>:MPHY:CRC:MAX.
*RST:
EQUal
TRIGger<m>:MPHY:CRC:MIN
<CRCMin>
Specifies the CRC bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<CRCMin> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:CRC:MAX <CRCMax>
Sets the end value of the CRC range if TRIGger<m>:MPHY:CRC:CONDition is set to
INRange or OORange.
Suffix:
<m>
.
1..3
Parameters:
<CRCMax> Numeric or string pattern, see Chapter 22.4.6, "Bit pattern
parameter", on page 1148. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:CREQ <REQ>
Specifies the CREQ bit pattern.
Suffix:
<m>
.
1..3
Parameters:
<REQ> ONE | ZERO | DC
Protocols