Acquisition and waveform setup
R&S
®
RTP
191User Manual 1337.9952.02 ─ 12
In addition to coupling, the signal can be filtered for high frequency rejection, see
Chapter 5.5.5, "Digital filter setup", on page 163.
"DC 1 MΩ"
Connection with 1 MΩ termination, passes both DC and AC compo-
nents of the signal.
"AC"
Connection with 1 MΩ termination through DC capacitor, removes DC
and very low-frequency components.
If AC coupling is set, the attenuation of passive probes has no effect,
and voltage is applied to the instrument with factor 1:1. Observe the
voltage limits, otherwise you can damage the instrument.
Remote command:
CHANnel<m>:COUPling on page 1199
5.6.11.2 Probe settings with connected adapter
Access: [Ch <n>] > "Channels" tab > "Probe Setup" button
In the "Probes Setup" tab, you can switch on the R&S ProbeMeter of the adapter, and
navigate to the adapter settings.
The R&S ProbeMeter of the adapter takes effect if the connected probe does not have
an own R&S ProbeMeter. Probes with an own R&S ProbeMeter use their own one.
Figure 5-17: Probe settings of R&S
RT-ZP10 connected via R&S
RT
‑
Z1M adapter, R&S ProbeMeter of
the adapter is available
5.6.11.3 R&S RT‑Z1M setup
Access: [Ch <n>] > "Channels" tab > "Probe Setup" button > "Adapter Setup" button >
"Setup" tab
In the "RT-Z1M Setup" tab, you can change the bandwidth limit, and start the self-align-
ment.
Probes