Acquisition and waveform setup
R&S
®
RTP
190User Manual 1337.9952.02 ─ 12
2. Connect the probe to the adapter.
The oscilloscope reads the probe-specific parameters through the adapter.
3. Set up the probe as usual, depending on the type of the connected probe.
4. In the "Adapter Setup", start the self-alignment of the adapter.
If the adapter is connected, additional settings are available in several dialog boxes:
● Channel settings with connected adapter............................................................. 190
● Probe settings with connected adapter.................................................................191
● R&S RT‑Z1M setup...............................................................................................191
● R&S RT‑Z1M attributes.........................................................................................192
5.6.11.1 Channel settings with connected adapter
Access: [Ch <n>] > "Channels" tab
In the "Channels" tab, the adapter is shown, and you can set the coupling and the
bandwidth limit here.
Bandwidth limit
Selects the bandwidth limit of the adapter. The specified bandwidth indicates the range
of frequencies that the instrument can acquire and display accurately with less than
3dB attenuation.
"Full"
At full bandwidth, all frequencies in the specified range are acquired
and displayed. Full bandwidth is used for most applications.
200 MHz, 20
MHz
Frequencies above the selected limit are removed to reduce noise at
different levels.
Remote command:
PROBe<m>:CONVerter:BANDwidth on page 1229
Coupling
Selects the coupling, which determines what part of the signal is used for waveform
analysis and triggering.
Probes