Remote control commands
R&S
®
RTP
1956User Manual 1337.9952.02 ─ 12
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:ENABle <CondEnabler>
Enables or disables the specific field within the defined frame to be filtered.
Suffix:
<m>
.
1..4
<n> *
<o> *
Parameters:
<CondEnabler> ON | OFF
*RST: OFF
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:IMAX
<IndexMax>
Sets the end value of an index range if BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:
IOPerator is set to INRange or RANGe.
Suffix:
<m>
.
1..4
<n> *
<o> *
Parameters:
<IndexMax> Range: 1 to 65534
Increment: 1
*RST: 65534
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:IMIN <IndexMin>
Specifies the index, or sets the start value of an index range.
Suffix:
<m>
.
1..4
<n> *
<o> *
Parameters:
<IndexMin> Range: 1 to 65534
Increment: 1
*RST: 1
BUS<m>:CMSB:FILTer:FRAMe<n>:FLD<o>:IOPerator <IndexOperator>
Sets the operator for the index in the selected field of the selected frame.
Suffix:
<m>
.
1..4