Remote control commands
R&S
®
RTP
1995User Manual 1337.9952.02 ─ 12
BUS<m>:MDIO:WCOunt?
Returns the word count for the selected serial bus, i.e. the number of words in the
present acquisition.
Suffix:
<m>
.
1..4
Return values:
<FrameCount> Range: 0 to 100000
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:MDIO:WORD<n>:DATA?
Returns the 16-bit payload data field content (in Clause 22 or Clause 45), or the 16-bit
address field content (in Clause 45, only) in the present acquisition of the selected
word and the selected serial bus. The most significant bit (MSB) is transmitted first.
Suffix:
<m>
.
1..4
<n> *
Return values:
<Data>
Usage: Query only
BUS<m>:MDIO:WORD<n>:PHYS?
Returns the 5-bit address field content (PHYAD/PRTAD) in the present acquisition of
the selected word and the selected serial bus.
Suffix:
<m>
.
1..4
<n> *
Return values:
<PhyAd> Range: 0 to 32
Increment: 1
*RST: 0
Usage: Query only
BUS<m>:MDIO:WORD<n>:REGI?
Returns the 5-bit register or device address field content (REGAD/DEVAD) in the pres-
ent acquisition of the selected word and the selected serial bus.