Remote control commands
R&S
®
RTP
1997User Manual 1337.9952.02 ─ 12
<n> *
Return values:
<State> OK | UNSYN | UNSYncronized | OPCO | TA_ERROR |
INComplete | SHORT | SHORt
OK
No error detected
UNSYN = UNSYncronized
UNSYncronized happens when bits are found, but they are not
correlated with any synchronization sequence. We don't know
what these bits are, but they are there. These bits receive a flag,
but they are not decoded.
OPCO
OPcode Error
TA_ERROR
Tirnaround time error
INComplete
Incomplete Frame
SHORT = SHORt
Length Error
*RST:
OK
Usage: Query only
BUS<m>:MDIO:WORD<n>:STOP?
Returns the stop time of the frame from the selected word within the search result.
Suffix:
<m>
.
1..4
<n> *
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:MDIO:WORD<n>:SYMBol?
Returns a textual translation (called Register Name) of the PHY or port address label in
the present acquisition of the selected word and the selected serial bus.
The translation is defined in the label list.
Suffix:
<m>
.
1..4
<n> *
Protocols