Remote control commands
R&S
®
RTP
2007User Manual 1337.9952.02 ─ 12
Parameters:
<ThresholdDminus> Range: -5 to 5
Increment: 0.01
*RST: 1.55
Default unit: V
BUS<m>:USB:DIFFerential:THReshold <ThresDiff>
Defines the threshold level for the differential signal in the USB high speed protocol.
Suffix:
<m>
.
1..4
Parameters:
<ThresDiff> Range: -2 to 2
Increment: 0.01
*RST: 0
Default unit: V
22.17.19.2
Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:USB:TYPE.............................................................................................. 2008
TRIGger<m>:USB:ACONdition......................................................................................2009
TRIGger<m>:USB:AMIN...............................................................................................2009
TRIGger<m>:USB:AMAX..............................................................................................2010
TRIGger<m>:USB:DATA...............................................................................................2010
TRIGger<m>:USB:DCONdition......................................................................................2010
TRIGger<m>:USB:DPOPerator..................................................................................... 2010
TRIGger<m>:USB:DPOSition........................................................................................2011
TRIGger<m>:USB:ECONdition......................................................................................2011
TRIGger<m>:USB:EMIN...............................................................................................2011
TRIGger<m>:USB:EMAX..............................................................................................2011
TRIGger<m>:USB:ERRC..............................................................................................2012
TRIGger<m>:USB:FCONdition......................................................................................2012
TRIGger<m>:USB:FMIN...............................................................................................2013
TRIGger<m>:USB:FMAX..............................................................................................2013
TRIGger<m>:USB:HAND..............................................................................................2013
TRIGger<m>:USB:PATT............................................................................................... 2013
TRIGger<m>:USB:PCONdition......................................................................................2014
TRIGger<m>:USB:PMIN...............................................................................................2014
TRIGger<m>:USB:PMAX..............................................................................................2014