Remote control commands
R&S
®
RTP
2244User Manual 1337.9952.02 ─ 12
TDRT:CALibration:OPEN[:EXECute]
Opens and loads the TDR/TDT calibration from the file specified with TDRT:
CALibration:OPEN:NAME.
Usage: Event
Asynchronous command
TDRT:TYPE <Type>
Selects the analysis type: TDR (time domain reflectometry), TDT (time domain trans-
mission), or TDR/TDT (both TDR and TDT).
Parameters:
<Type> TDR | TDRTdt | TDT
*RST: TDR
Usage: Asynchronous command
TDRT:CONFig <Configuration>
Selects the connection: single-ended or differential.
Parameters:
<Configuration> SINGleended | DIFFerential
*RST: SINGleended
Usage: Asynchronous command
TDRT:PULSe:OUT <Channel>
Sets the channel to which the "Out" connector of the pulse source is connected via cal-
ibration standard or DUT.
Parameters:
<Channel> C1W1 | C2W1 | C3W1 | C4W1
*RST: C2W1
Usage: Asynchronous command
TDRT:PULSe:REFerence <Channel>
Sets the channel to which the "Ref" connector of the pulse source is connected.
Parameters:
<Channel> C1W1 | C2W1 | C3W1 | C4W1
*RST: C1W1
Usage: Asynchronous command