Remote control commands
R&S
®
RTP
2250User Manual 1337.9952.02 ─ 12
Some of the commands in the following chapter are asynchronous. An overlapping or
asynchronous command does not automatically finish executing before the next com-
mand starts executing. If overlapping commands must be executed in a defined order,
e.g. to avoid wrong measurement results, they must be serviced sequentially.
To prevent an overlapping execution of commands, one of the commands *OPC,
*OPC? or *WAI can be used after the command or a command set.
For more information, see:
www.rohde-schwarz.com/rc-via-scpi, chapter "Command Sequence and Synchro-
nization"
Chapter 22.3, "Programming examples", on page 1126
IQ:STATe.....................................................................................................................2250
IQ:INSRate..................................................................................................................2250
CHANnel<m>:IQ:INPType.............................................................................................2250
CHANnel<m>:IQ:INPMode............................................................................................2251
IQ:RBWidth................................................................................................................. 2251
IQ:BWIDth?.................................................................................................................2251
IQ:SRATe....................................................................................................................2252
IQ:RLENgth.................................................................................................................2252
CHANnel<m>:IQ:CFRequency...................................................................................... 2252
CHANnel<m>:IQ:SBRF................................................................................................ 2253
CHANnel<m>:IQ:SBIF..................................................................................................2253
CHANnel<m>:IQ:WARNing[:STATe]?............................................................................. 2254
CHANnel<m>:IQ:WARNing:MESSage?..........................................................................2254
IQ:STATe <IQMode>
Activates the I/Q mode of the instrument.
Parameters:
<IQMode> ON | OFF
ON: I/Q mode
OFF: normal oscilloscope mode
*RST: OFF
Usage: Asynchronous command
IQ:INSRate <InputSampleRate>
Shows the sample rate of capturing the input data in I/Q mode: 20 Gsample/s.
If you use R&S RTP134 or R&S RTP164 instruments, you can change the sample rate
to 40 Gsample/s.
Usage:
Asynchronous command
CHANnel<m>:IQ:INPType <IQInputType>