Remote control commands
R&S
®
RTP
2285User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..2
Parameters:
<Position> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
ADVJitter<m>:DCOMposition:RESPlength <Length>
Set the step response length in unit intervals.
Suffix:
<m>
.
1..2
Parameters:
<Length> Range: 1 to 200
Increment: 1
*RST: 20
Default unit: UI
ADVJitter<m>:DCOMposition:ENP <State>
Enables the display of the periodic jitter components table.
Suffix:
<m>
.
1..2
Parameters:
<State> ON | OFF
*RST: ON
ADVJitter<m>:DCOMposition:NHORizontal <MaxNumber>
Sets the maximum number of periodic components in the horizontal direction that are
considered for the measurement. If the actual number of periodic components is higher
than the set value, they are considered for the random jitter component.
Suffix:
<m>
.
1..2
Parameters:
<MaxNumber> Range: 0 to 10
Increment: 1
*RST: 5
Jitter analysis and clock data recovery