Measurements
R&S
®
RTP
399User Manual 1337.9952.02 ─ 12
Note that the violation actions do not distinguish between a margin violation and a limit
violation. However, different icons are displayed in the result box.
For each action, you can define the event on which the action is initiated:
●
On violation
The action is initiated when the limits or margins are exceeded during the mea-
surement.
●
On successful completion
The action is initiated when a defined number of acquisitions has been captured,
and the limits or margins were not exceeded.
Independent of these actions, an icon is displayed in the result box, see Chapter 8.2.2,
"Measurement results", on page 336.
Beep
Generates a beep sound.
Remote command:
MEASurement<m>:ONViolation:BEEP on page 1427
Stop acq
Stops data acquisition on violation.
Remote command:
MEASurement<m>:ONViolation:ACQStop on page 1427
Save Wfm
Saves the waveform data to the file.
Remote command:
MEASurement<m>:ONViolation:WFMSave on page 1428
Trigger Out Pulse
Creates a pulse on the Trigger Out connector on limit violation.
When "Trigger Out Pulse" is used, the trigger control option "Enable trigger out" is dis-
abled. Thus, the trigger-out pulse is created only on limit violation but not when a trig-
ger occurs. The pulse is provided always with the minimum delay of 800 ns, the
"Delay" cannot be set.
See also: "Trigger out signal setup" on page 243.
Remote command:
MEASurement<m>:ONViolation:TRIGgerout on page 1428
Report
Creates and saves a report using the settings defined in "File" menu > "Report Setup".
Remote command:
MEASurement<m>:ONViolation:REPort on page 1428
Start Executable
Starts an external application. Tap "Config Executable" to set the application path and
parameters.
Automatic measurements