Mask testing
R&S
®
RTP
430User Manual 1337.9952.02 ─ 12
MTESt:SEGMent:REGion on page 1465
MTESt:SEGMent:COUNt? on page 1464
Configure
Opens a dialog to define the selected segment.
Definition of segment
The number of the selected segment is shown above the table. In the definition table,
the individual points of the selected mask segment are listed with exact horizontal and
vertical numerical coordinates. Here you can:
Insert a new point before the selected point.
Append a new point at the end of the list.
Remove the selected point from the list.
Change the x- and y-values of each point. To scale or move the complete segment,
use offset and factor values, see Rescale.
Remote command:
MTESt:SEGMent:POINt:ADD on page 1466
MTESt:SEGMent:POINt:REMove on page 1466
MTESt:SEGMent:POINt:INSert on page 1466
MTESt:SEGMent:POINt:X on page 1467
MTESt:SEGMent:POINt:Y on page 1467
MTESt:SEGMent:POINt:COUNt? on page 1466
Mask test settings