Protocol analysis
R&S
®
RTP
690User Manual 1337.9952.02 ─ 12
"Address
(Min)"
Defines the register address pattern for all operators that require one
pattern.
"Adress (Max)"
Defines the second address pattern that is required to specify a range
with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:ADDRess:CONDition on page 1819
TRIGger<m>:RFFE:ADDRess:MIN on page 1819
TRIGger<m>:RFFE:ADDRess:MAX on page 1820
Byte count
Sets the number of data frames to be read or written in the command sequence. The
setting is available for all "Extended Register" command sequences, which can trans-
fer more than one data frame.
"Condition"
Sets the operator to trigger on a specific byte count ("Equal" or "Not
equal") or a byte count range.
"Byte/Bit count
(Min)"
Defines the byte count for all operators that require one count setting.
"Byte/Bit count
(Max)"
Defines the second byte count value that is required to specify a
range with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:BCOunt:CONDition on page 1820
TRIGger<m>:RFFE:BCOunt:MIN on page 1820
TRIGger<m>:RFFE:BCOunt:MAX on page 1820
Data
Defines the data trigger condition. The data setup consists of the condition and one or
two data patterns.
To define on which data frames of the sequence you want to trigger, use the Index set-
tings.
"Condition"
Sets the operator to trigger on a specific data pattern ("Equal" or "Not
equal") or a data range.
"Data (Min)"
Defines the data pattern for all operators that require one pattern.
"Data (Max)"
Defines the second data pattern that is required to specify a range
with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:DATA:DCON on page 1821
TRIGger<m>:RFFE:DATA:DMIN on page 1821
TRIGger<m>:RFFE:DATA:DMAX on page 1821
Index
Defines which data frames are relevant for the trigger. The setting is available for all
"Extended Register" command sequences, which can transfer more than one data
frame.
The index allows you to check for a certain bit pattern at a certain position in the byte
sequence. Furthermore, you can trigger on the occurrence of a certain bit pattern
within a data range.
RFFE (option R&S
RTP-K40)