Protocol analysis
R&S
®
RTP
817User Manual 1337.9952.02 ─ 12
Payload
Defines, whether a payload check that meets specific conditions is executed or not.
"Condition"
Sets the operator "any" or "equal" that allows to trigger for payload
data at any position or at a specified position.
"Position"
Available only if "Condition" (TRIGger<m>:USB:DPOPerator) is set
to equal. Specifies the position in which a special data pattern is to be
triggered within the payload data packet.
"Data Condi-
tion"
Sets the operator ("equal" or "unequal", TRIGger<m>:USB:
DCONdition) to set a specific payload data pattern.
"Payload pat-
tern"
Specifies the payload data pattern that is to be triggered.
Remote command:
TRIGger<m>:USB:WPAY on page 2017
TRIGger<m>:USB:DPOPerator on page 2010
TRIGger<m>:USB:DPOSition on page 2011
TRIGger<m>:USB:DCONdition on page 2010
TRIGger<m>:USB:PATT on page 2013
Frame number
Defines, whether a frame number check that meets specific conditions is executed or
not.
"Condition"
Defining a specific frame number or a frame number range requires
to set the operator to one of the following: equal, not equal, less than,
less than or equal, greater than, greater than or equal, in range, out
of range.
"Min"
Specifies a frame number, or sets the start value of a frame number
range.
"Max"
Sets the end value of a frame number range if "Condition"
(TRIGger<m>:USB:FCONdition) is set to INRange or OORange.
Remote command:
TRIGger<m>:USB:WFRN on page 2017
TRIGger<m>:USB:FCONdition on page 2012
TRIGger<m>:USB:FMIN on page 2013
TRIGger<m>:USB:FMAX on page 2013
SC check
Defines, whether a Start / Complete SPLIT transaction check is executed or not.
Remote command:
TRIGger<m>:USB:WSTC on page 2018
TRIGger<m>:USB:STCO on page 2015
Port
Defines, whether a port check that meets specific conditions is executed or not.
USB (option R&S