Protocol analysis
R&S
®
RTP
818User Manual 1337.9952.02 ─ 12
"Condition"
Defining a specific port number or a port number range requires to
set the operator to one of the following: equal, not equal, less than,
less than or equal, greater than, greater than or equal, in range, out
of range.
"Min"
Specifies a port number, or sets the start value of a port number
range.
"Max"
Sets the end value of a port number range if "Condition"
(TRIGger<m>:USB:PCONdition) is set to INRange or OORange.
Remote command:
TRIGger<m>:USB:WPOR on page 2018
TRIGger<m>:USB:PCONdition on page 2014
TRIGger<m>:USB:PMIN on page 2014
TRIGger<m>:USB:PMAX on page 2014
SEU
Defines, whether an SEU check that meets specific conditions is executed or not. S
and E represent the Start and End of a start-split transaction, U represents the
reserved/Unused bit of a complete-split transaction. Permissible binary SEU values are
00, 01, 10, and 11.
"Condition"
Defining a specific SEU value or an SEU value range requires to set
the operator to one of the following: equal, not equal, less than, less
than or equal, greater than, greater than or equal, in range, out of
range.
"Min"
Specifies an SEU value, or sets the start value of an SEU value
range.
"Max"
Sets the end value of an SEU value range if "Condition"
(TRIGger<m>:USB:SCONdition) is set to INRange or OORange.
Remote command:
TRIGger<m>:USB:WSEU on page 2018
TRIGger<m>:USB:SCONdition on page 2014
TRIGger<m>:USB:SMIN on page 2015
TRIGger<m>:USB:SMAX on page 2015
ET
Defines, whether an Endpoint Type (ET) check that meets specific conditions is execu-
ted or not.
"Condition"
Defining a specific endpoint type or an ET range requires to set the
operator to one of the following: equal, not equal, less than, less than
or equal, greater than, greater than or equal, in range, out of range.
"Min"
Specifies an ET, or sets the start value of an ET range.
"Max"
Sets the end value of an ET range if "Condition" (TRIGger<m>:USB:
TCONdition) is set to INRange or OORange.
Remote command:
TRIGger<m>:USB:WETCheck on page 2017
TRIGger<m>:USB:TCONdition on page 2016