EasyManua.ls Logo

Rohde & Schwarz RTP044 - Page 942

Rohde & Schwarz RTP044
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mixed signal option (MSO, R&S RTP-B1)
R&S
®
RTP
942User Manual 1337.9952.02 ─ 12
Width
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
pulse width, respectively.
For the ranges "Within" and "Outside", the width defines the center of a range which is
defined by the limits "±Delta".
Remote command:
TRIGger<m>:PARallel:WIDTh:WIDTh on page 2211
±Delta
Defines a range around the given width value.
The combination "Range" = Within and "±Delta" = 0 triggers on pulses with a pulse
width that equals "Width".
The combination "Range" = Outside and "±Delta" = 0 means to trigger on pulse widths
"Width".
Remote command:
TRIGger<m>:PARallel:WIDTh:DELTa on page 2212
14.3.1.4 Timeout
The timeout trigger event checks if the trigger source signal stays above or below the
threshold voltage for a specified time lapse. In other words, the event occurs if the
state condition remains unchanged for the specified time.
You can use the timeout trigger on a single digital channel, or a logical combination of
digital channels.
Figure 14-3: Timeout trigger settings for trigger source = logical combination of digital channels
Range
Sets the state condition:
Trigger

Table of Contents

Other manuals for Rohde & Schwarz RTP044

Related product manuals