Pulse source and TDR/TDT analysis (option R&S RTP-B7/K130)
R&S
®
RTP
987User Manual 1337.9952.02 ─ 12
Enable
Switches all four pulse source outputs on and off.
Remote command:
PSRC[:STAT] on page 2241
Output level
Sets the low level of the output pulse. The high level is 0 V and cannot be changed.
Remote command:
PSRC:OUTPutlow on page 2241
Repetition rate
Sets the pulse frequency, the repetition rate of the pulse. Available values depend on
the selected clock mode.
Remote command:
PSRC:REPRate on page 2242
Duty cycle
Sets the duty cycle of the pulse in %, which is the ratio of the positive pulse width to
the period of the signal.
For repetition rates > 5 MHz, the value is fixed at 50 %.
Remote command:
PSRC:DUTYcycle on page 2242
Pulse source (option R&S
RTP-B7)

Table of Contents

Other manuals for Rohde & Schwarz RTP044