Routine Maintenance/Service Procedures
DSX™ System Service Manual 9-1
Chapter 9 Servicing the Incubator Module
9.1 Routine Maintenance/Service Procedures
Note: The Incubator Module is a service replaceable item. If a defect is observed,
the normal response is to replace the module. Replacement or repair of specific
items should be performed only on as needed basis if replacement of the module is
not possible and the service engineer has the appropriate components on hand.
After the module has been repaired, use the alignment plate to ensure that the
module is properly aligned as described in the Post-Service Checkout Protocol
section on page 3-6.
9.2 General Troubleshooting
Symptom Probable Cause Resolution
Plate motor fails
to work
There is a fault on
the IM Module Main
board.
Replace IM Module Main board.
Plate motor runs
at the incorrect
speed.
There is a fault on
the IM Module Main
board.
Replace IM Module Main board.
Plate carrier does
not move to the
home position
Plate home sensor is
faulty.
There is a fault on
the IM Module Main
board.
Replace faulty sensor.
Replace IM Module Main board.
Plate carrier
drives fully into
the IM Module.
Plate front sensor is
faulty.
IM Module Main
board is faulty.
Replace faulty sensor.
Replace IM Module Main board is faulty.