13.3.1 Dispense Bottle Assembly
The dispense bottle assembly includes a pump to deliver the fluid and a float to detect when
the fluid supply is exhausted. The dispense bottle assembly is normally replaced as a unit.
Instructions are also provided for replacing the dispense pump if needed.
To Replace the Dispense Bottle and Pump Assembly:
1 Remove the dispense bottle assembly (Figure 13-1) from the system.
2 Obtain the replacement dispense bottle assembly.
3 Place the replacement dispense bottle assembly on the system.
4 Calibrate the dispense pump using the Dynex Washer Calibration Software.
Figure 13-1 Dispense Bottle Assembly