Post-Service Checkout Protocol
DSX™ System Service Manual 3-11
3.5.5 Drawer Calibration
The drawer calibration locates all of the XYZ positions of the wells of all four plate
positions. It is performed if a new drawer module is installed, if the tip does not center
correctly, or if the tip does not go to the proper dispense height above the wells.
To Perform the Drawer Calibration:
1 Start Revelation, select Tools and then select Configure System. Select the
Coordinate Calibration tab of the Set-up DSX dialog box (Figure 3-2).
2 Select Calibrate Drawer. Follow the on-screen instructions to calibrate the
drawer.
3 The pipette spigot must enter the alignment hole vertically (directly overhead)
without any left-to-right or front-to-back side movement. Move the x-arm or y-
arm assemblies (if necessary) and push on the pipette only from the top.
4 Check for correct alignment several times. Hold the pipette spigot in the
alignment hole (without compressing the spigot spring) while selecting OK.