EasyManua.ls Logo

Dynex DSX - Manual Adjustment of the Module Calibration

Dynex DSX
358 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Post-Service Checkout Protocol
3-12 DSX™ System Service Manual
3.5.6 Manual Adjustment of the Module Calibration
Manual adjustment of a module may be required using a standard plate carrier if the unit
does not put and get the plates properly. Prior to starting this, the user should put a
standard carrier into the module that requires a manual adjustment.
1 Start Revelation, select Tools and then select Configure System. Select the
Coordinate Calibration tab of the Set-up DSX dialog box (Figure 3-2).
2 Select the module to manually adjust by left clicking on it with the mouse and
then select Manual Adjustment.
Note: The front left drawer is used as an example. The other modules are
adjusted in the same manner.
Figure 3-5 Front Left Drawer Manual Adjustment Dialog Box.
Note: The values in the Manual Adjustment Dialog Box for the three axes
represent approximately 0.01mm in travel. For example, the value reported
in Figure 3-5 for Z is equivalent to the step count (3566) multiplied by.0625
would provide the distance in millimeters from the Z home position. The
Y & Z step count multiplication factor is .0625 and the X is .0600. The home
position for the X is the farthest left position or 0, the Y is farthest inward
position or zero and the Z is the up most position or zero.

Table of Contents

Related product manuals