System Problems
DSX™ System Service Manual 4-7
Symptom Probable Cause Resolution
Error ____
loading DLL
____.
Program could not
find the required DLL
file
Re-install the program.
Error building
DCB.
Program could not
find a device.
Check the communication settings and
cables
Error importing
file nnn. Check
that the file is of
the correct type.
Program could not
open the file you
asked for
Check that the file is present and of the
correct type.
Error importing
SID file.
Program could not
open the file you
asked for.
Make sure that the file is correctly formatted.
See the File Format Conventions section in
user’s manual.
Error opening file
nnn.
Program could not
find the file you asked
for.
Recreate the file.
Error reading
assay.
Program could not
open the required
assay file because it
is not of the correct
type.
Change the Reader settings to the type that
was used to create the file, or try using the
File menu, Import facility to import it.
Error setting
comm state.
Communication
between systems lost
Check the communication settings and cable.
File ___ is not the
correct version.
File ___ is not a
Revelation format
file.
The file you have tried
to open is not a
standard format used
by this version of
Revelation
Error setting comm state.
Try importing the file using the File menu
Import option.
Hardware
link...Hardware
not present.
Cable accidentally
disconnected
Check the communication settings and
cables.
Installation
Failure.
During installation,
Revelation tried to
use a file that was
being used by another
Windows program.
Close down all other programs and run the
Setup program again. If Microsoft Office is
running, remove its icon from the Startup
group and reload Windows. You can replace
Office in the Startup group once Revelation is
loaded.