System Problems
4-8 DSX™ System Service Manual
Symptom Probable Cause Resolution
Insufficient
Memory.
A memory conflict has
occurred.
If running a complex test, try to simplify it. If
you are running other applications, close
them.
Internal
Application Error.
A fault has occurred in
the program
Try repeating the command or reboot the
program
Invalid byte size.
Check the communication settings and
cables.
Invalid data file
format.
The data file you are
trying to open is not a
standard format used
by Revelation.
To open the file, try using the File menu,
Import facility.
Module Not
Loaded.
This assay contains
operations that are
not available for your
Reader model, or you
have selected the
wrong Reader.
Remove the unsupported operation, set
Revelation up for the correct module or check
to see if the correct module is installed.
No timers
available.
Revelation can not
run correctly because
of memory conflicts
Close down some of the other applications
and try again; if the error persists try resetting
the PC.
OLE 2.0
Initialization
failed.
A fault has occurred in
the program
Try repeating the command or reboot the
program
Unable to
allocate queues.
Check the communication settings and
cables