GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 3 - Setup Instructions 3-41
Table 3-15 Scan Assistant
Preset Parameter Description
Auto Selection Off, Category, or Description.
• Off. The Scan Assistant selection on the Patient screen is completely manual. It will say
“None” when starting a new patient and you can make a selection manually if desired.
• Category. Scan Assistant uses the combination of exam category (Abd, OB, etc.) and the
currently-selected user to automatically select a Scan Assistant program. It chooses the
same program that was used the last time this combination of exam category and user.
The user can manually override this auto selection.
• Description. Scan Assistant uses the combination of the exam description (often auto fills
in if the patient was selected from the worklist) and the currently selected user to
automatically select a Scan Assistant program. It will choose the same program that was
used the last time this combination of exam description and user was selected. The user
can manually override this auto selection. If the exam description is blank then it will do
the auto selection based on Category as described in the previous paragraph.
Always Use Doppler Cursor Use the Doppler Cursor when you activate Scan Assistant.