GE HEALTHCAREDRAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-143
Section 8-37
Replacement of DCWD ASSY
8-37-1 DCWD ASSY
8-37-1-1 Manpower
One person, 15 minutes
8-37-1-2 Tools
Philips screw driver
8-37-1-3 8-32-1-3 Removal procedure
1.) Remove PIL ASSY, refer to 8-16-4-2 Removal Procedure - Separating CWD Board from PIL.
2.) Remove three screws from back of CW connector board
Figure 8-166 CW connector board
3.) Locate CW connector on PIL board
Figure 8-167 PIL board
8-37-1-4 Installation Procedure
Parts to be installed in reverse order on removal
8-37-1-5 Functional Check
Refer to Section 4-4 Functional Diag Test List