GE HEALTHCARERAFT LOGIQ™ S7 EXPERT/PRO
DIRECTION 5460683, REVISION 3 DRAFT (JUNE 25, 2014) SERVICE MANUAL
8-130 Section 8-31 - Replacement of the Monitor Arm
Section 8-31
Replacement of the Monitor Arm
NOTE: Some of the photos resembles consoles other than LOGIQ™ S7 Expert/Pro, but Monitor Arm is
common part as Polaris Platform and therefore, service procedure is the same.
8-31-1 Manpower
One person, 5 minutes.
8-31-2 Tools
Phillips screwdriver, Nipper, Allen.
8-31-3 Removal Procedure
1.) Unscrew 1 screw, then remove the 3rd arm cover. Refer to the figure below.
2.) Remove the 2nd Axis Cover. Refer to the figure below.
Figure 8-150 Removing the 3rd arm cover
Figure 8-151 Removing the 2nd Axis Cover